27 JULY TLD (flight hardware review) 1. fly 6 dosimeters + 1 calibration light source. 2. the light source will be launched in the reader, + should be left in the reader (to avoid debris problems) between data collection periods 3. a data collection period will consist of 2 readings of light source (1st) 2 readings of each dosimeter 1 final reading of light source (to recheck calibration) 4. the battery should be on for a couple of minutes before the readings....procedure: turn on reader, go collect dosimeters, then start readings NOTE: battery life is 100 hrs (~4 days) 5. to take a reading: - check BATT-ON, (check READY lt) - insert dosimeter (position 1) - turn to position 2 (check MEAS lt) (starts ~40 sec measurement cycle) - read + record display - take 2nd reading by turning to pos 1, and back to 2 - check BATT-off